![]() |
|
Figure 1: The Select Experiment, PeakForce TUNA mode, window.
![]() |
|
|
Figure 2: TUNA control parameters in the Scan Parameter list
|
PF-TUNA Parameter | Use with PF- TUNA |
---|---|
Current sens |
Selects the gain of the TUNA sensor:
|
DC Test Bias | Provides an additional bias for the purpose of testing the TUNA sensor. |
DC Sample Bias | Selectable bias to chuck/sample. |
Table 1: PF-TUNA Control parameters
Figure 3: NanoScope gain switch messages
Figure 4: The PeakForce TUNA Gain switch
Ensure that you reset the parameters to their default values (12) when using the standard air tapping or TR holder.
www.bruker.com | Bruker Corporation |
www.brukerafmprobes.com | 112 Robin Hill Rd. |
nanoscaleworld.bruker-axs.com/nanoscaleworld/ | Santa Barbara, CA 93117 |
Customer Support: (800) 873-9750 | |
Copyright 2010, 2011. All Rights Reserved. |