You are here: Installation and Maintenance Guide > Calibration

Calibration

This section provides detailed instructions for the fine calibration of the Dimension Icon SPM. In general the scanner’s calibration should be checked every three months. The main change in calibration will be the loss of piezo sensitivity in the first year of use. This is an expected behavior; loss of 10–15% of the initial scan size is common.

Bruker employs a software-guided calibration procedure for all microscopes. The procedural details of how calibration is executed using NanoScope software are beyond the scope of this User Guide and include proprietary methods exclusive to Bruker. The calibration is summarized in four basic steps outlined below.

  1. Orthogonality Adjustments: The orthogonality adjustment compensates for a possible non-square scan pattern.
  2. Scanner Linearization: Adjust parameters for correcting the capacitive nature of the piezo material and force the XY motion of the piezo to move at a constant velocity.
  3. XY Calibration: Designate the software conversion values using one or both of the following methods.

    -OR-

  4. Calibrate Z: Calibrate the Z scan tube to within 2%. A separate calibration “look-up” table stores the conversion values for converting the control voltages applied to the Z tube. These values are determined by scanning over a sample with a known depth.

Bruker recommends that you adhere to the Calibration schedule shown in table 1. After initial installation, perform the Fine-tuning for XY Calibration per the following time schedule. If you find that the calibration measurements are more than 10% off at any point during Fine-Tuning Calibration, stop and perform the full XY calibration routine (Capture Calibration). For most applications it is sufficient to perform the Z calibration with the same frequency. For critical height measurements Bruker recommends monthly Z calibration.

Calibration Routine Time Frame Frequency
Fine-Tuning Calibration (or Full XY Calibration if required) First Year Every 3 months
Subsequent Years Every 6 months
Z Calibration (for General Applications) First Year Every 3 months
Subsequent Years Every 6 months
Critical Height Measurements All Years Monthly

Table 1: Recommended calibration schedule

NOTE: If using scan sizes of 5 μm or smaller, Bruker recommends calibrating the scanner for small scan sizes. Contact Bruker for further instructions.

Depending on your application or the amount of time your system has been in use, you may need to run through one or more of the various Calibration Procedures to ensure optimal functioning of your Dimension Icon.

 

www.bruker.com Bruker Corporation
www.brukerafmprobes.com 112 Robin Hill Rd.
nanoscaleworld.bruker-axs.com/nanoscaleworld/ Santa Barbara, CA 93117
   
  Customer Support: (800) 873-9750
  Copyright 2010, 2011. All Rights Reserved.