You are here: Installation and Maintenance Guide > Installation > Installing the Dimension Icon System > Connect the Stage Controller

Connect the Stage Controller

  1. Connect cables to the Dimension Stage Controller. See the table below and figure 1 for a list of cables, connection information and connection locations.
NOTE: The Dimension Icon cables are bundled through the cable clamp provided if the system includes an acoustic hood.
Figure Reference Cable Part Number Box Function
1 AC Power 466-000-004 Power Strip Dimension controller power
P6/P7 Vacuum Hose Assembly 860-000-012 Dimension Icon Vacuum and air supply
P1 Serial Cable (6') 464-000-024 Computer Serial communication with computer
P3 Serial Cable   NanoScope V Controller Crash protection
P4 37-pin Cable   Dimension Icon Motion control
P5 37-pin Cable   Dimension Icon Motion control

Figure 1: Dimension Stage Controller - Rear View

  1. To set illumination, rotate the manual light control switch on the lower right front of the Dimension stage controller to On for manual control or Off for computer control:

Figure 2: Dimension Stage Controller - Front View

NOTE: Each illumination system may vary in intensity. Adjust to preferred levels once the system is operational—for computer adjustment there is an Illumination parameter in the Align and Navigate windows.
  1. Remove the temporary cover from the vacuum/air ports.
  2. Toggle the chuck vacuum switch, located on the front of the microscope unit, to Off:

 

www.bruker.com Bruker Corporation
www.brukerafmprobes.com 112 Robin Hill Rd.
nanoscaleworld.bruker-axs.com/nanoscaleworld/ Santa Barbara, CA 93117
   
  Customer Support: (800) 873-9750
  Copyright 2010, 2011. All Rights Reserved.