Troubleshooting
The Dimension Icon system is very reliable and generally will not need adjustment once properly calibrated, nor is it likely to develop a problem due to use. In most cases any problems that occur can be linked to some external event.
Some typical examples of such events are:
- Lighting storms that cause power transients
- Fluid spills on electrical components
- Improper installation of the air table when relocating the instrument to a new region
- After relocating the instrument to a new room, the data may exhibit noise. The room itself may have excessive floor vibration or acoustic noise.
- Improper signal cable installation when relocating the instrument
- Installing a new board other than what is needed to operate the instrument via the computer can cause the software to not load correctly or a hardware conflict within the computer system.