If any of the three power Diagnostics LEDs (i.e., +5 V, +15 V, –15 V) on the front panel, shown in Figure 1, fails to illuminate when the system is turned On, a fuse may be blown.
Figure 1: Power Supply LEDs on the Front of the NanoScope V Controller
To replace the fuse, complete the following:
|
|
|
| Fuse | 100 V | 120 V | 240 V |
|---|---|---|---|
| 1 | 2.5 A | 2.5 A | 2.0 A |
| 2 | 800 mA | 800 mA | 400 mA |
| 3 | 800 mA | 800 mA | 400 mA |
| 4 | 10 A | 10 A | 6.3 A |
| ALL FUSES: 250 V SLOW BLOW | |||
NOTE: If the problem persists, contact Bruker to determine which system components to return for repair.
|