It is important to choose a probe that can cause enough deformation of the sample and still retain high force sensitivity. Therefore cantilever stiffness should be selected based on the sample stiffness. Bruker's recommendations are shown in Table 1.
Sample Modulus (E) | Probe | Nominal Spring Constant (k) |
---|---|---|
1 MPa < E < 20 MPa | ScanAsyst-Air | 0.5 N/m |
5 MPa < E < 500 MPa | Tap150A, P/N MPP-12120-10 | 5 N/m |
200 MPa < E < 2000 MPa | Tap300A (RTESPA), P/N MPP-11120-10 | 40 N/m |
1 GPa < E < 20 GPa | Tap525A, P/N MPP-13120-10 | 200 N/m |
10 GPa < E < 100 GPa | DNISP-HS | 350 N/m |
To reduce optical interference, probes should be coated on their back side.