Extended TUNA SSRM tip holder
This tip holder is a bit different from the other holders, it is a bit
thicker. This can cause self oscillation in the closed loop mode of
the scanner piezos.
There is a way to reduce the feedback gain in the closed loop, here
is info how to access and set this.
Please remember to set back the X and Y gains to
12.00 when the session is finished.
Access scanner settings via Calibrate - Scanner - X-Y...
Change the X and Y gains to 10.00 for operation with the Extended TUNA
SSRM tip holder.
Please remember to set back the X and Y gains to
12.00 when the session is finished.
Anders Liljeborg
Nanostructure Physics, KTH.