Usage rules and instructions for the
AJA Ion miller Albanova (139)

Responsible: Erik Holmgren

 

General Rules:

Usage instructions:

Sample holder Loading/Unloading:

  1. Vent chamber by flipping the switch on the main rack labeled VENT/PUMP to the off position
  2. Wait for the pressure to reach 100 Torr (1*10+2 on gauge) then open manual vent valve fully and loosen the screw on the door. Do not open the manual vent valve before or the turbo can be damaged
  3. Wait for the pressure to reach 750 Torr, then open the door
  4. Screw the sample holder into/out-of the chamber-arm, for easy access make sure the arm is tilted to -90 degrees. The holder should be finger-tight, do not over-tighten or the carbon backing plate can be damaged
  5. Shut off the manual vent valve, close the door and tighten screw
  6. Flip the switch on the main rack labeled VENT/PUMP to the on position
  7. Wait until the turbo is at full speed (820 Hz) and the pressure is below 5*10-6

Etching:

  1. Make sure the turbo is at full speed (820 Hz) and the pressure is below 5*10-6
  2. Turn on the Kaufman power supply in the main rack, wait for start-up routine to finish (about 10 seconds)
  3. Make sure the Kaufman power supply is in remote and auto mode (indicators on right-hand-side of power supply)
  4. If necessary start the PhaseIIJ software and log in, password is on the screen
  5. Click "Run Process" and select the correct process, then click "Open"
  6. Wait for the process to finish
  7. Make sure the required 10 minute flush at 50 sccm Ar starts after the process finishes
  8. Turn off the Kaufman power supply in the main rack

Sample mounting instructions:

Full 100mm wafers may be mounted directly on the sample holder.

All samples smaller than a full 100mm wafer must be secured on the chip-adapter Note: If the screws securing the clamps are becoming hard to tighten/loosen, do not force them in/out contact the tool responsible

Making recipes:

Each process is comprised of several steps (referred to as "Layers" in the software). To make a new process click "Create Process" on the main screen to open the process creator, then select the layers to add to the process, click add and then save the new process. EACH PROCESS MUST END WITH THE "10 min Ar Purge" LAYER.

To make a new layer click "Create Layers" on the main screen to open the layer creator. Turn on the desired functions (for example Ar gas flow or substrate rotation) and enter the desired parameters in the white boxes, then click save and enter a new for the layer. The newly made layer should now appear in the list of layers in the process creator.

The ion gun will not operate at arbitrary parameters and care has to be taken to avoid damaging the ion optics.
To choose etching parameters, first choose a beam voltage (lower voltage gives gentler etch), then set acceleration voltage to at least 20% of the beam voltage, then set the desired beam current below the maximum indicated in the table below;
Beam voltage [V] Minimum acceleration voltage [V] (must be <20% of V_beam) Maximum beam current [mA]
400 80 55
600 120 120
800 160 200
1000 200 290 (special care required)

Some etch rates are indicated in the tables below;

Beam voltage 600 V, Acceleration Voltage 120, Beam current 120
Material Etch rate [nm/min]
Si, SiO2, maN14xx, Cr (Edwards), LN ∼17
Ta, W, TaN ∼5
Cr (AJA1) ∼10

Beam voltage 400 V, Acceleration Voltage 80, Beam current 55
Material Etch rate [nm/min]
Si, SiO2, maN14xx ∼5
Ta, W, TaN ∼4
Cr, LN, HfO2 ∼7
Ag, Au, Cu ∼30
Mo, Al, Si-N ∼9

The etch rate depends on the angle of incidence, the rate decrease as roughly the cosine of the angle. As the angle is varied a mask will cast a shadow which limits the minimum separation of objects, the shadow length is equal to the tangent of the angle of incidence multiplied with the mask thickness. The etching rate relative to 0 degrees angle of incidence and the resulting normalized shadow cast by your mask is indicated in the table below;

Angle of incidence, a Relative etching rate (cos(a)) Normalized shadow (tan(a))
0 1 0
15 0.97 0.27
30 0.87 0.58
45 0.71 1
60 0.5 1.73
75 0.26 3.73
90 0 inf

End of user instructions


Webmaster Albanova Nanolab