According to recent discussions in cleanroom meetings and in order to
avoid inconsistent etch rates, each user of Cryo RIE machine is
supposed to run a cleaning recipe after they are done with the tool:
Before your RIE session:
If there is a fused silica carrier wafer, put it in the wafer
container labeled post Cl2 cleaning recipe, which can be
found next to
the load lock.
After your RIE session:
- Remove your sample and carrier wafer
- Load the carrier wafer labeled post Cl2 cleaning
recipe
- Run cleaning recipe:
- If you have been using Cl2-based etching, run the recipe
named 1: postCl2 clean
- Otherwise, run the recipe 2: O2 Clean
These recipes are the first ones in the recipes list
-
Log the cleaning recipe in the log file on the computer
-
You can leave the tool now, since the next users is supposed to unload the cleaning carrier wafer before their using the machine.
|