You may tune the SCM sensor for maximum sensitivity using Sweep commands:
![]() |
|
Figure 1: SCM Sensor Tuning Curve Parameter Settings, Tip Engaged
NOTE: Sensitivity is optimized when the Capacitance Sensor Frequency is set to the frequency corresponding to the highest slope of the tuning (sensor resonance) curve.
Tuning the SCM sensor frequency is complicated by the change in resonant frequency with engage. In a sweep width of approximately 890 MHz to 1050 MHz, the peak may shift 50–100 MHz, due to increasing capacitance as the probe tip approaches the sample. Two tuning curves are displayed in Figure 2 (withdrawn) and Figure 3 (engaged), illustrating the resonance shift with tip/sample proximity. |
Figure 2: Capacitive Sensor Frequency tune (tip withdrawn)
Figure 3: Setting the Capacitance Sensor Frequency (Tip Engaged)
A better way to perform this tune is to Engage then start the Generic Sweep by raising the probe tip 100–150 nm above the sample surface (see Figure 4) and then adjusting the VCO (Voltage Controlled Oscillator) on the side of the SCM Application Module using a non-metallic screwdriver (supplied) to maximize the slope before setting the center frequency. The probe tip will return to the surface when you Exit the Generic Sweep function. |
Figure 4: Raise the Probe Tip 150 nm when starting the Generic Sweep
www.bruker.com | Bruker Corporation |
www.brukerafmprobes.com | 112 Robin Hill Rd. |
nanoscaleworld.bruker-axs.com/nanoscaleworld/ | Santa Barbara, CA 93117 |
Customer Support: (800) 873-9750 | |
Copyright 2010, 2011. All Rights Reserved. |