You are here: Experiment Guide > Imaging Modes > Electrochemistry > LakeShore 331S Front Panel Control

LakeShore 331S Front Panel Control

Front Panel Displays

The temperature controller is pre-programmed with default settings (see Default Temperature Controller Settings) that influence the liquid crystal display (LCD) readings. The default settings are appropriate for standard use of the temperature controller with the Dimension Icon Electrochemical Cell; no modifications are needed.

The symbols displayed by the LCD are:

Figure 1: Front of LakeShore 331 Temperature Controller

Typically, after initial power-up, the LCD is treated as four equal-sized subpanels (see Figure 1). To change what is displayed in one of the subpanels:

  1. Press the Display Format button (under the LCD, top row, button “4”). “Display Location 1” appears written in the lower half of the LCD.
  2. Use the up/down arrow buttons to pick among:

    Press Enter.

  3. Use the up/down arrow buttons to pick among:

    Press Enter.

  4. Use the up/down arrow buttons to pick a source among:

    Press Enter.

NOTE: If Setpoint is entered for display location 3, only the first three source options are available. If Heater out is entered for display location 4, there are no additional options to enter.
NOTE: To display a calculated temperature in location 4, see Math Functions and Temperature Calibration.

The six front panel light emitting diode (LED) “annunciators” record system status in normal operation. In normal operation, only the first light, Control A, is on.

NOTE: The IEEE-488 interface is not necessary for normal Dimension Icon Electrochemistry Cell function.
NOTE: Alarms have not been programmed as a part of Dimension Icon Electrochemistry Cell function, but can be used if desired.

Basic Front Panel Operation

The full set of front panel functions is listed below. However, only a few of the front panel buttons are regularly used:

To enter a temperature setpoint:

  1. Press Setpoint.
  2. Enter a number (via the numbered dual function buttons in the left half of the front panel buttons).
  3. Press Enter.

To turn on the heater:

  1. Press Heater range.
  2. Toggle the arrow buttons to select High.
  3. Press Enter.

To turn off the heater:

NOTE: These basic functions can also be controlled with the NanoScope software.

Front Panel Button Functions

There are 20 front panel function buttons (see Figure 1). Here they are listed from top-to-bottom and left-to-right and described briefly. See Default Temperature Controller Settings for the default values given to these parameters initially, and refer to the LakeShore 331 User’s Manual for more detail about each function.

NOTE: Press and hold Escape to reset instrument and assign LakeShore default values (which are different from Bruker default values (see Default Temperature Controller Settings).
NOTE: The Dimension Icon Electrochemistry Cell employs control loop 1 exclusively.

 

www.bruker.com Bruker Corporation
www.brukerafmprobes.com 112 Robin Hill Rd.
nanoscaleworld.bruker-axs.com/nanoscaleworld/ Santa Barbara, CA 93117
   
  Customer Support: (800) 873-9750
  Copyright 2010, 2011. All Rights Reserved.