You are here: System Overview > TAC Heater/Cooler > Operation > Experimentation Procedures

Experimentation Procedures

Prior to imaging samples at non-ambient temperatures, inspect the cooling lines for air bubbles. Bubbles can destabilize temperature control of the sample. A few minutes of pumping prior to heating is sufficient to eliminate bubbles from the coolant. This is always advisable after the system has been off for several hours. Verify that fluid flows at the output of the reservoir.

In a typical experiment, a sample is examined at room temperature before its temperature is changed to a target value.

CAUTION: Disengage the probe from the sample prior to heating to avoid unwanted contact as the sample and heater components expand with temperature. Only small (5–10°C) temperature increases can be performed safely without tip withdrawal. By the same logic, large (> 30°C) temperature rises require further removal of the probe tip from the sample.

NOTE: In imaging polymers at elevated temperatures in TappingMode, drive amplitude may be required to overcome increased sample stickiness.

Measurements at temperatures up to 75°C can be performed without powering the probe heater. However, when operating at higher temperatures, it is advised to use the Tip Heater to avoid condensation of moisture or the deposition of volatile sample components on the cantilever. These contaminants can destabilize the cantilever resonance and reduce the optical reflectivity to the laser beam.

Stable imaging in TappingMode is the main purpose of the Tip Heater. Due to difficulties in measuring probe temperature, its heating is regulated by a voltage applied to the heater, which is installed in direct contact with the probe substrate. An increase of 1 volt raises substrate temperature approximately 10°C. It has been demonstrated that the application of ≤  7 V is sufficient for stable imaging in the range of 75–140°C. The applied voltage needed gradually increases for imaging at higher temperatures. However, long-term operation at voltages above 15 V may diminish the life of the piezostack-activated probe tapping oscillator.

NOTE: Use of the probe heater may lead to a slight change in the average level of the probe, causing a shift of the reflected laser beam away from the center of the position-sensitive photodetector. Therefore, during high temperature operation, watch the differential (A-B) signal from the vertical segments of the detector and adjust manually as needed. Also, the heater can influence the coupling between the piezostack and the cantilever, requiring a different drive amplitude to generate, for instance, 2 V RMS oscillation of the cantilever at a different temperature. Again, it is worth monitoring the cantilever amplitude sweep (Cantilever Tune) and retuning the cantilever to a target oscillation amplitude after the amplitude vs frequency curve has stabilized following a temperature adjustment.

Gas Purging

The heated TappingMode (air) probe holder in combination with the silicone rubber seal provides the option to control the atmosphere around the heated sample by purging with inert, dry, non-corrosive gases (e.g., nitrogen, argon, helium, etc).

Because of the possibility for material oxidation at high temperature, the gas-tight heater/sample chamber can be purged of oxygen with an inert gas (e.g. nitrogen, argon, helium, etc.). The small volume of the sample enclosure allows purging to be accomplished in 1–3 minutes at a rate of 5–10 ml/min. Gas replacement verification is easily checked using helium as a substitute for air because the resonant frequency of the cantilever in helium is slightly raised, while its quality factor increases markedly.

The flow meter calibration curve, shown in Figure 1, shows flow rate, in ml/min vs. the scale markings on the flow meter.

Figure 1: Flow meter calibration curve

Purging the sample chamber with inert gas leads to an increase in heat consumption and thus can reduce the maximum attainable sample temperature or increase the minimum attainable sample temperature.

Dry nitrogen gas purging, at 50–500 ml/min, may not be required for sub-ambient operation in air, but is required for temperatures near and below 0°C. Flow rates less than 50 ml/min. are generally too low to prevent ice formation, while flow rates greater than 500 ml/min will heat the sample or cause noise. Dry nitrogen from a high-pressure gas cylinder may not be dry enough to prevent ice formation on the sample surface at low temperatures. In this case, Bruker recommends the use of dry nitrogen from a gas vent on a low-pressure liquid nitrogen container. Minimize the nitrogen gas flow to prevent sample heating by the relatively warm gas while preventing ice formation on the sample surface. Bruker has found several suitable sources for liquid nitrogen storage:

MVE Bio-Medical Systems, Chart Industries, Inc.
3505 County Road 42 West
Burnsville, MN 55306-3803 USA
Web: www.chart-ind.com (www.chartbiomed.com)
E-mail: storagesystems@chart-ind-com
United States: 800-400-4683 Fax: 952-882-5191
Worldwide: 952-882-5090 Fax: 952-882-5008
Chart Europe GmbH: +49 (0)212-700 570, Fax: +49 (0)212-700 577
Chart Asia Inc.: 65-838-5209, Fax: 65-235-3680
Product Information:
Storage Vessel Model: # CryoCyl 35 (35 Ltr, 22 psi) or CryoCyl 50 (50 Ltr, 22 psi)
Transfer Hose: 4 or 6 Foot Transfer Hose
Gas Vent Connection: Check with vendor.
CryoFab
540 Michigan Avenue
P.O. Box 485
Kenilworth, NJ 07033 USA
Web: www.cryofab.com
E-mail: sales@cryofab.com
Product Information:
Storage Vessel Model: # CLPB25 (25 Ltr, 20 psi) or # CLPB50 (50 Ltr, 20 psi).
Transfer Hose: 4 or 6 foot transfer hose, non-insulated or vacuum insulated, can be purchased.
Gas Vent Connection: 3/8 NPT Male Connector on gas vent.
Wessington Cryogenics
Building 9, Philadelphia Complex
Houghton-le-Spring, Tyne & Wear
DH4 4UG, ENGLAND
Web: www.wessingtoncrygenics.com
E-mail: info@wessingtoncryogenics.co.uk
Product Information:
Storage Vessel Model: # PV-30 (30 Ltr, 1.5 bar) or # PV-60 (60 Ltr, 1.5 bar).
Transfer Hose: May be required for filling vessel.
Gas Vent Connection: Check with vendor.

Thermoelectric Element Cooling

Room temperature water, with the coolant reservoir, is sufficient for heating and for cooling to approximately –20°C.

Reaching the lowest specified operating temperature requires the use of ice water as the coolant directly from the ice bucket. Remove the silicone tubes from the reservoir and insert them into the ice water bucket. Reaching the lowest specified operating temperature also requires use of no sample puck or the smallest sample puck (6 mm).

Each time the pump system is started the user should inspect for leaks, particularly near the base of the scanner. Periodically monitor the cooling system for leaks during operation and replace leaky tubing.

When the pump is first started, the pulse damper (see Cooling System) will take a few minutes to fill before cooling water starts to flow back into the reservoir. Furthermore, coolant flow will not stop immediately when the pump is turned off because the coolant inside the pulse damper will continue to flow for several minutes.

Because the polyurethane tubing is transparent, you can visually monitor the filling process. Verify that coolant is flowing freely into the reservoir. You should see a drip/stream from the scanner side of the reservoir cap.

The Heater/Cooler element should not be enabled until the cooling system is completely filled.

Temperature Calibration

Bruker sets the heater and temperature sensors, which do not normally require recalibration. The temperature measured and controlled by the heater controller is that of the heating/cooling element, not the sample. For the most accurate sample temperature measurement, measure the sample temperature independently using an optional temperature sensor.

 

www.bruker.com Bruker Corporation
www.brukerafmprobes.com 112 Robin Hill Rd.
nanoscaleworld.bruker-axs.com/nanoscaleworld/ Santa Barbara, CA 93117
   
  Customer Support: (800) 873-9750
  Copyright 2010, 2011. All Rights Reserved.