You are here: System Overview > TAC Heater/Cooler > Operation > Heater/Cooler Software Control

Heater/Cooler Software Control

If you have a NanoScope V Controller and NanoScope 7.30 or later software, the Thermal Applications Controller can be controlled using the NanoScope software, as described below. Earlier NanoScope or NanoDrive software requires front panel control, discussed in TAC Panel Control.

Configure the NanoScope Software

  1. Launch the NanoScope software by clicking on the desktop microscope icon.
  2. Set the microscope configuration by selecting Tools > Select Microscope... to open the Microscope Select window:

  1. Click the Edit button to open the Equipment window, then click the Advanced button to show all the equipment options:

  1. Select the TAC Controller in the Temp Controller field:

  1. Click OK to apply this selection.
  2. Configure the serial ports by clicking the Serial... button in the Microscope Select window. This will open the Serial Port Configuration window:

  1. Highlight appropriate port and click the Edit button to open the Edit Port Setup window:

  1. Use the drop-down menu to select TAC Temp Control in the Equipment Type field and click OK to close the Edit Port Setup window for that port.
  2. Click Done to close the Serial Port Configuration window.
  3. Click OK to close the Microscope Select window.

Software Control

Sample and tip heater temperatures can be controlled through the Temperature Control panel, available in either the Scan view:

Temp Panel Parameters:

Auto Z-Stepper Motor Control

Z-drift can be a significant problem for heating (and cooling) due to thermal expansion. The purpose of the Auto Z-Step function is to keep the Z-center position from becoming extended or retracted, as a result of z-drift.

To activate Auto Z-Step:

  1. Engage on the sample.
  2. In the Temperature Control parameter list, change the Auto Z-Step parameter to Enabled. The Auto Step Motor window will appear.
  3. Configure the Auto Step Motor and click the Start button.
  4. To turn off the Auto Step Motor function, click the Stop button.

With Auto Z-Step enabled in the Temperature Control panel and started via the window, NanoScope software automatically steps the z-stage up or down whenever the z-center voltage drifts too far towards the fully retracted or fully extended position. These two z-positions are user defined by Retracted Threshold and Extended Threshold, shown in Figure 1. When the z-center voltage is less than Retracted Threshold, then the z-stage is stepped "up" (by the Motor Step Size) until the z-center is at the Target Z-Center.

The Withdraw Threshold, for the retracted side only, protects against feedback errors, which cause the Z piezo to fully retract. If the Z-center is less than the Retracted Threshold, the probe is automatically withdrawn.

Figure 1: Auto Step Motor window

Auto Z-Step logic is shown in Figure 2 with the Auto Step Motor functions and symbols defined below.

Figure 2: Auto Z-Step logic

Parameter Settings

Retracted Voltage Threshold (RVT)

  • Range: Withdraw Threshold +20 V to Target Z-Center –20 V
  • Default Setting: –210 V

Extended Voltage Threshold (EVT)

  • Range: Target Z-Center +20 V to ZMax (hybrid head: 110V, other heads 220 V).
  • Default Setting: 60 V

Target Z-Center (TZC)

  • Default Setting: –50 V (hybrid head) 0 V (other heads)

Withdraw Threshold (WT)

  • Range: Withdraw Threshold –20 V to ZMax (hybrid head: 110V, other heads: 220 V).
  • Default Setting: –210 V

Motor Step Size

  • Range: 0 μm to 0.5 μm
  • Default Setting: 0.4 μm

Start

  • Starts the Auto Step Motor function

Stop

  • Stops the Auto Step Motor function

Exit

  • Closes the Auto Step Motor window

Target ± tolerance (T)

  • Not user adjustable or visible in the Auto Step Motor window
  • Default Setting: 10 V

Table 1: Parameters and Setting Ranges and Defaults for the Auto Step Motor Window

Short Withdraw

The purpose of the short withdraw function is to provide you with a safe means to change temperature, adjust the detector and feedback settings, tune the cantilever tune, etc. in close proximity to the surface so that the temperature of the tip and sample remain stable and the gas seal is maintained.

The Short Withdraw function is available only when the tip is engaged on the sample surface. To execute a Short Withdraw, select Microscope > Short Withdraw from the NanoScope menu.

To change the short withdraw distance (the factory default is 20 μm):

  1. Click Microscope > Engage Settings to open the Engage Parameters window:

Figure 3: Engage Parameters window

  1. Modify the Short withdraw distance (range is 0 to 999 μm). The distance should be large enough to get off the sample surface but close enough to maintain a consistent temperature. Values between 20 and 50 μm are suggested.
  2. Click OK.

Gain Reset

Due to the additional mass of the Heater/Cooler TappingMode Probe Holder, some systems may oscillate following engage. In this situation, the software will automatically reset the x and y integral gain values to protect the scanner. An error log dialog window will appear to inform the user. See Closed Loop XY Calibration for more information.

 

www.bruker.com Bruker Corporation
www.brukerafmprobes.com 112 Robin Hill Rd.
nanoscaleworld.bruker-axs.com/nanoscaleworld/ Santa Barbara, CA 93117
   
  Customer Support: (800) 873-9750
  Copyright 2010, 2011. All Rights Reserved.