See Table 1 for characterization of the Dimension Heater/Cooler capabilities and Table 3 for conditions associated with its controller.
| Specification | |||
|---|---|---|---|
|
Parameter |
High Temperature Heater |
Heater/Cooler |
60°C Heater (Bio Heater) |
|
Maximum Heater Temperature (Air)1 |
250°C |
100°C |
62.5°C |
|
Maximum Heater Temperature (Fluid)1 |
|
50°C |
50°C |
|
Minimum Heater Temperature (Air)1 |
Ambient |
Ambient | |
|
Minimum Heater Temperature (Fluid)1 |
Ambient |
< 4°C |
Ambient |
|
Pressure |
Ambient |
Ambient |
Ambient |
|
Resolution |
0.1°C |
0.1°C |
0.1°C |
|
Accuracy |
3.0% |
3.0% |
3.0% |
|
Temperature Drift1 |
< 0.5°C |
< 0.5°C |
< 0.5°C |
|
Overshoot1 |
< 1.0°C4 |
< 1.0°C typical < 2.5°C max5 |
< 1.0°C6 |
|
Heating Rate Nominal1 |
> 10°C/min |
> 20°C/min |
1–2°C/min |
|
Cooling Rate Nominal1 |
n/a |
> 20°C/min |
1–2°C/min |
Table 1: Specifications for the Dimension Heater/Cooler
|
Zones |
Stage Type |
Medium |
Temperature Range |
Gain |
||
|
|
|
|
|
P |
I |
D |
|
1 |
High T Heat |
Air |
Ambient to 250°C |
102 |
80 |
3 |
|
3 |
Low T Heat |
Air |
Ambient to 62.5°C |
78 |
85 |
3 |
|
5 |
ThermoElectric Stage |
Air |
50°C to 100°C |
102 |
42 |
3 |
|
6 |
ThermoElectric Stage |
Air |
35°C to 50°C |
126 |
42 |
3 |
|
7 |
ThermoElectric Stage |
Air |
25°C to 35°C |
150 |
42 |
3 |
|
8 |
ThermoElectric Stage |
Air |
15°C to 25°C |
150 |
42 |
3 |
|
9 |
ThermoElectric Stage |
Air |
0°C to 15°C |
102 |
25 |
3 |
|
10 |
ThermoElectric Stage |
Air |
–40°C to 0°C |
82 |
20 |
3 |
|
11 |
Heat Stage |
Fluid |
Ambient to 250°C |
102 |
80 |
3 |
|
13 |
Low T Heat |
Fluid |
Ambient to 62.5°C |
78 |
70 |
13 |
|
15 |
ThermoElectric Stage |
Fluid |
50°C to 100°C |
60 |
40 |
0 |
|
16 |
ThermoElectric Stage |
Fluid |
35°C to 50°C |
60 |
40 |
0 |
|
17 |
ThermoElectric Stage |
Fluid |
25°C to 35°C |
30 |
25 |
0 |
|
18 |
ThermoElectric Stage |
Fluid |
15°C to 25°C |
30 |
25 |
0 |
|
19 |
ThermoElectric Stage |
Fluid |
0°C to 15°C |
30 |
25 |
0 |
|
20 |
ThermoElectric Stage |
Fluid |
–40°C to 0°C |
30 |
25 |
0 |
|
Parameter |
Specification |
|
Range of Environmental Conditions |
5°–55°C, Relative Humidity 5%–90%, non-condensing |
|
Range of Environmental Conditions |
5°–70°C |
|
Rated Voltage |
100, 120, 230 or 240 volts AC (V~) |
|
Maximum Rated Current |
1.6 A @ 100, 120 V, 50/60 Hz |
Table 3: Specifications for High Temperature Heater Controller Electronics
Table 4 through Table 8 list TAC default settings. Settings that differ from Watlow CLS200 default values are shown in bold. See the Watlow CLS200 Series User’s Guide for detailed setup instructions.
|
Global Parameters |
TAC Default Value (All Loops) |
|
Load Setup from Job |
1 |
|
Save Setup to Job |
1 |
|
Job Select Digital Inputs |
None |
|
Job Select Digital Inputs Active |
Low |
|
Output Override Digital Input |
None |
|
Override Digital Input Active |
Low |
|
Startup Alarm Delay |
0 min. |
|
Keyboard Lock Status |
Off |
|
Power Up Output Status |
Off |
|
Process Power Digital Input |
None |
|
Controller Address |
1 |
|
Communications Baud Rate |
19200 |
|
Communications Protocol |
Mod (Modbus RTU) |
|
AC Line Frequency |
60 Hz |
|
Digital Output Polarity on Alarm |
Low |
|
EPROM information: |
Varies* |
Table 4: Setup Global Parameters Default Values
|
Setup Loop Input |
TAC Default Value (vs. Loop) |
||
|
Loop = 01 (TE) |
Loop = 02 (HT) |
Loop = 03, 04, 05 |
|
|---|---|---|---|
|
Input Type |
K T/C |
K T/C |
Skip |
|
Loop Name |
TE |
HT |
03 to 05 |
|
Input Units |
°C |
°C |
°F |
|
Input Reading Offset |
0°C |
0°C |
0°F |
|
Reversed T/C Detection |
Off |
Off |
Off |
|
Input Filter |
Varies* |
Varies* |
3 Scans |
Table 5: Setup Loop Input Default Values
|
Setup Loop Control Parameters |
TAC Default Value (vs. Loop) |
||
|
Loop = 01 (TE) |
Loop = 02 (HT) |
Loop = 03, 04, 05 |
|
|---|---|---|---|
|
Heat Control PB (Proportional Band) |
Varies* |
Varies* |
50°F |
|
Heat Control TI (Term Integral) |
Varies* |
Varies* |
180 Sec./R |
|
Heat Control TD (Term Derivative) |
Varies* |
Varies* |
0 Sec. |
|
Heat Control Output Filter |
3 |
3 |
3 |
|
Cool Control Proportional Band |
Varies* |
Varies* |
50°F |
|
Cool Control TI (Term Integral) |
Varies* |
Varies* |
60 Sec./R |
|
Cool Control TD (Term Derivative) |
Varies* |
Varies* |
0 Sec. |
|
Cool Control Output Filter |
3 |
3 |
3 |
|
Spread |
0.5 |
0.5 |
0.5 |
|
Restore PID Digital Input |
None |
None |
None |
Table 6: Setup Loop Control Parameters Default Values
|
Setup Loop Control Outputs |
TAC Default Value (vs. Loop) |
||
|
Loop = 01 (TE) |
Loop = 02 (HT) |
Loop = 03, 04, 05 |
|
|---|---|---|---|
|
Heat Control Output |
Enabled* |
Enabled* |
Enabled |
|
Heat Output Type |
DZC |
DZC |
TP |
|
Heat Output Action |
Reverse |
Reverse |
Reverse |
|
Heat Output Limit |
100% |
100% |
100% |
|
Heat Output Limit Time |
Continuous |
Continuous |
Continuous |
|
Sensor Fail Heat Output |
0% |
0% |
0% |
|
Heat T/C Break Output Average |
Off |
Off |
Off |
|
Heat Output |
Linear |
Linear |
Linear |
|
Cool Control Output |
Enabled* |
Disabled |
Disabled |
|
Cool Output Type |
DZC |
– |
– |
|
Cool Output Action |
Direct |
– |
– |
|
Cool Output Limit |
100% |
– |
– |
|
Cool Output Limit Time |
Continuous |
– |
– |
|
Sensor Fail Cool Output |
0% |
– |
– |
|
Cool T/C Break Output Average |
Off |
– |
– |
|
Cool Output |
Linear |
– |
– |
Table 7: Setup Loop Outputs Default Values
|
Setup Loop Alarms |
TAC Default Value (vs. Loop) |
||
|
Loop = 01 (TE) |
Loop = 02 (HT) |
Loop = 03, 04, 05 |
|
|---|---|---|---|
|
High Process Alarm Setpoint |
Varies* |
Varies* |
1000 |
|
High Process Alarm Type |
Alarm |
Alarm |
Off |
|
High Process Alarm Output Number |
None |
None |
None |
|
Deviation Alarm Value |
0.5°C |
0.5°C |
0.5°C |
|
High Deviation Alarm Type |
Off |
Off |
Off |
|
High Deviation Alarm Output Number |
None |
None |
None |
|
Low Deviation Alarm Type |
Off |
Off |
Off |
|
Low Deviation Alarm Output Number |
None |
None |
None |
|
Low Process Alarm Setpoint |
–45°C |
0 |
0 |
|
Low Process Alarm Type |
Alarm |
Off |
Off |
|
Low Process Alarm Output Number |
None |
None |
None |
|
Alarm Deadband |
5.0°C |
5.0°C |
0.2°C |
|
Alarm Delay |
0 seconds |
0 seconds |
0 seconds |
Table 8: Setup Loop Alarms Default Values