It is generally not difficult to obtain accurate XY calibration references. It is, however, much more difficult to obtain accurate Z-axis results. Z axis calibration is very sample-dependent. It is difficult to control Z piezo dynamics because, during scanning the X- and Y-axes move at a constant rate while the Z axis does not. Furthermore, offsets affect the piezo over a period of minutes.
The height and Z sensor signals require separate calibrations. Because the Z sensor is linear, only one calibration constant is required. However, the height sensitivity (Z Scan Sens) varies as a function offset, so multiple calibration points are necessary. We use Z Offset = 0 and Z offset = ± 50 V.
The silicon calibration references distributed by Bruker have specified (on the calibration container) vertical features accurate to within ± 3 percent. The calibration reference is referred to throughout the examples provided in this section. If you require greater accuracy, you must select an appropriate calibration standard, and a closed loop scanner employed with a Bruker Dimension Series microscope.
The remaining parameters in the Z Calibration panel are detailed below.
Z Min: The lowest voltage allowed to be applied to the negative terminal of the Z-piezoelectric actuator.
Z Max: The largest voltage allowed to be applied to the positive terminal of the Z-piezoelectric actuator.
CAUTION: Z Min and Z Max are set at the factory or during installation of the scanner and are based on the safe operating range of the Z-axis piezoelectric tube. Lowering Z Min or raising Z Max to outside the factory-set range can result in irreparable damage to the scanner. Consult with Bruker first if you believe Z Min or Z Max should be adjusted.
(Z–(–Z)) Min: Twice the absolute valure of Z Min, computed automatically, requiring Z Max ≥ Z Min +10 V.
(Z+(–Z)) Max: Twice the absolute value of Z Max, computed automatically, requiring Z Max ≥ Z Min +10 V.
Several additional parameters are used to calibrate the Z-axis for closed loop operation:
Z Sensor Sens: The sensitivity, in nm/V, of the optical Z-axis sensor when Z Sensor Gain Factor = 1. Determine this by measuring a calibration standard with the Z sensor channel. See Z Sensor Sensitivity.
Z Sensor Gain Factor: During the engage, the Z piezo is ramped to find the Z piezo range. The sensor gain is adjusted to set a Z sensor range that is:
Z CL Range Factor: The range of the Z sensor that is used in Z closed loop mode (ramping). Increasing this factor increases the chance that part of the ramp may reach the end of the scanner range.
www.bruker.com | Bruker Corporation |
www.brukerafmprobes.com | 112 Robin Hill Rd. |
nanoscaleworld.bruker-axs.com/nanoscaleworld/ | Santa Barbara, CA 93117 |
Customer Support: (800) 873-9750 | |
Copyright 2010, 2011. All Rights Reserved. |