You are here: Installation and Maintenance Guide > Installation > Installing the Dimension Icon System > Connect the NanoScope V Controller to a New Native SPM

Connect the NanoScope V Controller to a New Native SPM

Connections to the microscope are made through connectors on the front of the NanoScope V Controller. See Table 1 for descriptions of the main front panel connectors.

Plug the cables from your new Dimension SPM or BioScope Catalyst E-box into the NanoScope V Controller through the connectors labeled Microscope (High Voltage/Digital and Low Voltage) at the right front of the controller (see Figure 1).

If you have a Closed Loop head (CLH) (XY, Hybrid XYZ, Icon, Dimension FastScan, BioScope II or BioScope Catalyst), connect its preamp or the Dimension Icon E-box to the Closed Loop Scanner connector at the right front of the controller (see Figure 1).

If you have a PicoAngler, connect it to the Angler connector at the right front of the controller (see Figure 1).

For a cabling overview, refer to Dimension Icon: Computer to Controller Cable Connections .

 

Connector Label Connector Type Purpose
High Voltage/Digital 36-pin combo D-sub, female Piezo drive/digital control between controller and SPM
Low Voltage 78-pin DB, male Differential low voltage analog/digital (LVDS) signals between controller and SPM
Angler 25-pin DB, female Communicates with PicoAngler
Computer (back panel) 68-pin micro-SCSI, female Data/control between NanoScope V Controller and computer (DSI)
Expansion (back panel) USB type B Data/control between NanoScope V-PI Controller and computer (DSI)
Closed Loop Scanner 14-pin Lemo, female Input/output to the XYZ Closed Loop Scanner Head or Closed loop BioScope II/Catalyst head and XY scan stage
Digital I/O BNC, female TTL compatible synchronization signals (i.e. EOL and EOF)
High Frequency/ Analog BNC, female Used as either a lock-in input or for HSDC. Sample rate is 50 MS/s.
Low Frequency Analog BNC, female Used for external measurement. Can bring up as a data type in NanoScope software.

Table 1: NanoScope V Controller, Main Connections

Figure 1: Microscope Connectors on the Front of the NanoScope V Controller

 

www.bruker.com Bruker Corporation
www.brukerafmprobes.com 112 Robin Hill Rd.
nanoscaleworld.bruker-axs.com/nanoscaleworld/ Santa Barbara, CA 93117
   
  Customer Support: (800) 873-9750