You are here: Installation and Maintenance Guide > Installation > Installing the Dimension Icon System > Unpacking

Unpacking

The NanoScope V Controller unit ships in a single crate. Additional computer and SPM system components are crated separately.

CAUTION: Do not store the equipment outside, even in a dry weather location.

Provide the proper environmental conditions (see Table 1) for system operation and storage.

Equipment Use Condition Requirement
Operation Temperature 15ºC (59ºF) to 35ºC (95ºF)
Operation Relative Humidity 45% to 80%, non-condensing
Storage Temperature 10ºC (50ºF) to 50ºC (122ºF)
Storage Relative Humidity 35% to 90%, non-condensing

Table 1: Environmental Requirements

Carefully unpack the controller from its crate.

CAUTION: Handle sensitive electronics with care. Avoid dropping or bumping the controller, especially once outside the shipping crate.

The unit measures 60 cm (23.6 inches) long by 45 cm (17.7 inches) wide by 19 cm (7.5 inches) tall, including its four feet.

Caution/Attention/Vorsicht: The NanoScope V Controller weighs 18 kg (40 lbs) so requires two people to move safely.

CAUTION: Each controller is shipped equipped for a customer-specified AC input voltage. Input power requirements are listed as a function of AC input voltage in a table on the back of the controller (and duplicated below in Table 2). Verify appropriate fuses are installed for the power supplied. A fuse rating table appears on the back of the controller and is duplicated later (see Table 1 in Fuses).
Voltage (VAC) Current (A) Frequency (Hz)
100 5.2 50 – 60
120 4.3 50 – 60
240 2.2 50 – 60

Table 2: AC Input Power Requirements

 

www.bruker.com Bruker Corporation
www.brukerafmprobes.com 112 Robin Hill Rd.
nanoscaleworld.bruker-axs.com/nanoscaleworld/ Santa Barbara, CA 93117
   
  Customer Support: (800) 873-9750
  Copyright 2010, 2011. All Rights Reserved.