The NanoScope V Controller unit ships in a single crate. Additional computer and SPM system components are crated separately.
Provide the proper environmental conditions (see Table 1) for system operation and storage.
Equipment Use | Condition | Requirement |
---|---|---|
Operation | Temperature | 15ºC (59ºF) to 35ºC (95ºF) |
Operation | Relative Humidity | 45% to 80%, non-condensing |
Storage | Temperature | 10ºC (50ºF) to 50ºC (122ºF) |
Storage | Relative Humidity | 35% to 90%, non-condensing |
Table 1: Environmental Requirements
Carefully unpack the controller from its crate.
The unit measures 60 cm (23.6 inches) long by 45 cm (17.7 inches) wide by 19 cm (7.5 inches) tall, including its four feet.
Voltage (VAC) | Current (A) | Frequency (Hz) |
---|---|---|
100 | 5.2 | 50 – 60 |
120 | 4.3 | 50 – 60 |
240 | 2.2 | 50 – 60 |
Table 2: AC Input Power Requirements