The Thermal Tune method, based on Measuring Thermal Noise, provides an automated and quick determination of cantilever spring constant.
A correction factor should be applied to this calibration. When we determine the deflection sensitivity, we exert a static load at the end of the cantilever, but when we measure the thermal energy the cantilever is oscillating at resonance and is constrained at only one end rather than two. The cantilever bends differently in the two situations and thus reflects the laser beam of the optical lever differently. See D. Sarid, “Scanning Force Microscopy With Applications to Electric, Magnetic and Atomic Forces,” 2nd Edition, Oxford University Press, New York 1994, pp 5–6 and pp 10–11. When you calibrate the deflection sensitivity using a static force, you underestimate the amplitude of dynamic motion by approximately 8% (for a simple beam). This correction factor is applied using the Deflection Sensitivity Correction parameter in the Thermal Tune window. |
||
![]() |
|
|
![]() |
|
|
Figure 1: The Thermal Tune window
Typically the markers are located roughly where the spectrum rises from the noise floor. Precise placement is unnecessary; the fit is insensitive to the minimal power contributed from these frequencies far from the natural resonance (see right “shoulder” of waveform in Figure 1). You may exclude a larger portion of a shoulder of the waveform from the fit bandwidth to ignore a noise spike. Experiment with repeated fits of the same acquired thermal tune data to become familiar with its sensitivity to bandwidth and choice of model.
NOTE: While the goal of setting the bandwidth for the curve fit was to achieve the highest signal-to-noise ratio by excluding noise, in setting the bandwidth for spring constant calculation from the (noise-free) fit curve the objective is to include all frequencies that contribute power to the spectrum. For instance, if you left out part of a shoulder in performing the curve fit, include it in determining the spring constant.
Clicking Yes copies the calculated spring constant in the “Ramp” list in the PicoForce view. |
www.bruker.com | Bruker Corporation |
www.brukerafmprobes.com | 112 Robin Hill Rd. |
nanoscaleworld.bruker-axs.com/nanoscaleworld/ | Santa Barbara, CA 93117 |
Customer Support: (800) 873-9750 | |
Copyright 2010, 2011. All Rights Reserved. |