TAC Panel Control

With a NanoScope V Controller and NanoScope 7.30 or later software, the Thermal Applications Controller can be controlled by NanoScope software (see Heater/Cooler Software Control). Earlier NanoScope or NanoDrive software requires front panel control, discussed below.

Front Panel Control

Set and control the sample temperature by regulating the heater or heater/cooler element and tip heater voltage with the Thermal Applications Controller (TAC). This controller includes an over-temperature alarm mode that turns the heater off when the reservoir temperature exceeds a factory setpoint. The TAC front panel consists of the following components, explained below:

CAUTION:
Watlow™ Controller Panel
Manual Setpoint Control Enable/Disable switch
Manual Setpoint Control knob
Stage Enable/Disable/Reset switch
Over-Temperature Alarm Shutdown
Manual Tip Voltage
Left-Side Display
Right-Side Display

Rear Panel Functions

Controller Power: The power plug, switch and main fuse are located on the back of the control box.

The Serial Port is connected to a serial port connection, usually COM4, of the PC controlling the Dimension Icon. The serial port is used only for NanoScope (software) control of the TAC. Connection is unnecessary for manual control.

Figure 4: Thermal Applications Controller (Rear Panel)

CAUTION: Assure that the voltage configuration is set to the appropriate regional power supply. An incorrect setting may cause equipment damage (i.e., a blown fuse).

If necessary, change the operating voltage by opening the fuse box cover, as shown in Figure 5, with a small screwdriver and removing and rotating the fuse holder.

Figure 5: Fuse Holder

 