With a NanoScope V Controller and NanoScope 7.30 or later software, the Thermal Applications Controller can be controlled by NanoScope software
Set and control the sample temperature by regulating the heater or heater/cooler element and tip heater voltage with the Thermal Applications Controller (TAC). This controller includes an over-temperature alarm mode that turns the heater off when the reservoir temperature exceeds a factory setpoint. The TAC front panel consists of the following components, explained below:
Controller Power: The power plug, switch and main fuse are located on the back of the control box.
The Serial Port is connected to a serial port connection, usually COM4, of the PC controlling the Dimension Icon. The serial port is used only for NanoScope (software) control of the TAC. Connection is unnecessary for manual control.
Figure 4: Thermal Applications Controller (Rear Panel)
CAUTION: Assure that the voltage configuration is set to the appropriate regional power supply. An incorrect setting may cause equipment damage (i.e., a blown fuse).
If necessary, change the operating voltage by opening the fuse box cover, as shown in Figure 5, with a small screwdriver and removing and rotating the fuse holder.