NanoScope V Front Panel
The NanoScope V Controller, shown in Figure 1, has the following items on the front panel:
- Controller status LCD.
- Power Supply LEDs (top 3) — indicates the power supply status. These are discussed in System Initialization.
- Diagnostic LED (“active/flashing”) — indicates the FPGA is loaded.
- Connection to the (optional) PicoAngler.
- Connections for Dimension XY, Dimension Hybrid XYZ, Dimension Icon or Dimension FastScan AFM closed loop heads or the BioScope Catalyst microscope.
- Digital I/O BNCs:
- Inputs (In 1/2) — Pulse counter inputs.
- Output 1 (Out 1) — End of Line: TTL pulse at the end of each scan line.
- Output 2 (Out 2) — End of Frame: TTL pulse at the end of each scan frame.
- High Frequency Analog:
- Lock-In 1/2 Inputs — Analog inputs for two high frequency lock-ins accessible via NanoScope software. The sample rate is 50 MHz giving an effective bandwidth of 1 kHz to 5MHz.
- DDS 1/2 Outputs — Analog output of the reference signals for two high frequency lock-ins accessible via NanoScope software (1 kHz - 5 MHz).
- Low Frequency Analog
- Lock-In 3 Input — Analog input for a low frequency lock-in accessible via NanoScope software. The sample rate is 500 kHz giving an effective bandwidth of 5 Hz to 50 kHz. If this lock-in is not used, the BNC connector is available as a third General I/O Input (see below).
- DDS 3 Output — Analog output of the reference signal for the low frequency lock-in accessible via NanoScope software.
- General I/O Inputs 1/2 — Analog inputs that can be read as data channels in the NanoScope software. Continuous data collection is done at a sample rate of 100 kHz. “Burst” capture (data collection for a limited duration of time) can be done at a sample rate of 500 kHz via the High Speed Data Capture function in the NanoScope software.
- General I/O Outputs 1/2 — Analog outputs that can be used to monitor various signals within the controller selectable via Nanoscope software. The update rate is 100 kHz (40 kHz in ramp mode).
Figure 1: NanoScope V Front Panel
The Microscope Connectors on the right of the NanoScope V Controller front panel are detailed in Connect the NanoScope V Controller to a New Native SPM.