Figure 1: The Engage Parameters window showing the Actuated Probe Engage panel.
The Actuated Probe Engage panel serves the same purpose in the dual control loop Actuated Probe Mode as the Tapping Engage panel does for traditional control loop TappingMode in earlier NanoScope software. That is, each contain parameters that control the tip/sample engage process.
To open the Actuated Probe Engage panel (see Figure 1), select Microscope > Engage Settings > Actuated Probe Engage.
parameter | Description |
---|---|
TM engage gain |
The normal (scanner) piezoelectric actuator integral gain value used during Actuated Probe (and TappingMode) engage. Once the surface is engaged, the value reverts to that of Feedback> TM engage gain. Range and Settings:
Hints for Optimizing TM Engage Gain
|
Tapping Engage Fast I Gain |
The fast (cantilever) piezoelectric actuator integral gain value used during Actuated Probe engage in TappingMode. Once the surface is engaged, the value reverts to that of Feedback> Fast Integral gain. Range and Settings:
|
Tapping Engage Fast P Gain |
The fast (cantilever) piezoelectric actuator proportional gain value used during Actuated Probe engage in TappingMode. Once the surface is engaged, the value reverts to that of Feedback> Fast Proportional gain. Range and Settings:
|
Contact Engage Fast I Gain |
The fast (cantilever) piezoelectric actuator integral gain value used during Actuated Probe engage in Contact Mode. Once the surface is engaged, the value reverts to that of Feedback> Fast Integral gain. Range and Settings:
|
Contact Engage Fast P Gain |
The fast (cantilever) piezoelectric actuator proportional gain value used during Actuated Probe engage in Contact Mode. Once the surface is engaged, the value reverts to that of Feedback> Fast Proportional gain. Range and Settings:
|
CAUTION: If the probe is allowed to engage by actuation of the (fast) cantilever piezo alone, with the scanner tube gains set to zero, the probe tip may be driven into the sample when the scanner tube gains are raised to nonzero values. The Minimum Scanning Gain parameters protect against this.
|
|
Tapping Minimum Scanning I Gain |
The minimum integral gain value the normal Z-actuator may have during TappingMode engage. This value overrides any lower setting of Feedback > Integral gain, including remaining in effect after engage. Range and Settings:
|
Tapping Minimum Scanning P Gain |
The minimum proportional gain value the normal Z-actuator may have during TappingMode engage. This value overrides any lower setting of Feedback > Proportional gain, including remaining in effect after engage. Range and Settings:
|
Contact Minimum Scanning I Gain |
The minimum integral gain value the normal Z-actuator may have during Contact Mode engage. This value overrides any lower setting of Feedback > Integral gain, including remaining in effect after engage. Range and Settings:
|
Contact Minimum Scanning P Gain |
The minimum proportional gain value the normal Z-actuator may have during Contact Mode engage. This value overrides any lower setting of Feedback > Proportional gain, including remaining in effect after engage. Range and Settings:
|
Tapping Maximum Scan Rate |
The maximum scan rate value allowed before TappingMode engage completes. This value overrides any higher setting of Scan Controls/Scan Rate made before engage completes. If an override occurs, the maximum remains in effect after engage until changed in Scan Controls/Scan Rate. Range and Settings:
|
Contact Maximum Scan Rate |
The maximum scan rate value allowed before Contact Mode engage completes. This value overrides any higher setting of Scan > Scan Rate made before engage completes. If an override occurs, the maximum remains in effect after engage until changed in Scan> Scan Rate. Range and Settings:
|
Determine Contact Gain |
Set to Enabled to obtain accurate height data with both slow and fast control loops active. After Determine Contact Gain has been Enabled, prior to engage, the cantilever piezoelectric actuator is ramped over ±10V drive amplitude while monitoring deflection. This contact gain calibration records the resultant deflection per applied drive signal for future use. Range and Settings:
|
Check for active tip |
Automatically verifies a functioning Active Tip probe before each engage operation. Range and Settings:
|
www.bruker.com | Bruker Corporation |
www.brukerafmprobes.com | 112 Robin Hill Rd. |
nanoscaleworld.bruker-axs.com/nanoscaleworld/ | Santa Barbara, CA 93117 |
Customer Support: (800) 873-9750 | |
Copyright 2010, 2011. All Rights Reserved. |