Next, you set scan parameters and scan the sample.
|
Scan Size: |
1 µm |
Aspect Ratio: |
1.00 |
Scan Angle: |
0.00 ° |
Scan Rate: |
1 Hz |
|
Aspect Ratio: |
8.00 (4.00 if Lines is 1024; 1.00 if Lines is 5120) and you want square pixels |
Samples/Line: |
5120 |
Lines: |
640, 1280 or 5120 |
NOTE: You can use the mouse to adjust the value in many parameter fields. Click on the value and drag the mouse left to decrease the value or right to increase the value.
The Aspect Ratio controls the X:Y ratio of the pixels in the displayed image. Because there is an 8:1 ratio between 5120 samples/line and 640 lines, using an Aspect Ratio of 8 causes the pixels displayed in the image to be square. NOTE: Some microscopes, e.g. Dimension Icon-PI, are limited to 1 K (1024) lines and 1 K points per line.
|
SPM feedback: |
Amplitude |
Integral gain: |
0.5 |
Proportional gain: |
5.0 |
|
Integral Gain: |
2.0 |
Proportional Gain: |
5.0 |
Deflection Setpoint: |
0 V (vertical deflection = –2 V (before engage) |
|
|
|
|
NOTE: You can zoom in on the scan image by selecting the Zoom button below the image. Then, use your mouse to drag a box outline over the area you want to zoom in on. Click Offset to offset the center position of your scan.
|
|
|
Preparing a Dimension Series AFM for a Realtime Scan
Analyzing an Image With Section Analysis
www.bruker.com | Bruker Corporation |
www.brukerafmprobes.com | 112 Robin Hill Rd. |
nanoscaleworld.bruker-axs.com/nanoscaleworld/ | Santa Barbara, CA 93117 |
Customer Support: (800) 873-9750 | |
Copyright 2010, 2011. All Rights Reserved. |